Journal of Histochemistry and Cytochemistry Priciples for Free Access to Science
  Search:   
    >> Advanced Search

Guidelines | Subscriptions | About | exPRESS - Current - Archive | Business Information | Contact
This Article
Right arrow Full Text (PDF)
Right arrow Alert me when this article is cited
Right arrow Alert me if a correction is posted
Services
Right arrow Similar articles in this journal
Right arrow Similar articles in PubMed
Right arrow Alert me to new issues of the journal
Right arrow Download to citation manager
Right arrow reprints & permissions
Citing Articles
Right arrow Citing Articles via Google Scholar
Google Scholar
Right arrow Articles by Alabaster, O.
Right arrow Articles by Chen, R. F.
Right arrow Search for Related Content
PubMed
Right arrow PubMed Citation
Right arrow Articles by Alabaster, O.
Right arrow Articles by Chen, R. F.
Social Bookmarking
 Add to CiteULike   Add to Complore   Add to Connotea   Add to Del.icio.us   Add to Digg   Add to Reddit   Add to Technorati  
What's this?

Electrolytic degradation of DNA fluorochromes during flow cytometric measurement of electronic cell volume

O Alabaster, DL Glaubiger, VT Hamilton, SA Bentley, SE Shackney, KS Skramstad and RF Chen

Changes in flow cytometric measurement of DNA content can result from electrolytic chemical degradation of mithramycin, ethidium bromide, and propidium iodide during simultaneous measurement of electronic cell volume. Bench electrolysis also degrades these fluorochromes without changing the quantum yields, even when they are complexed to DNA. In the flow cytometer, electrolytic production of chlorine at the anode is the probable cause of this degradation, since exposure of these fluorochromes to chlorine gas produces the same effect. It is therefore advisable to measure the DNA content distribution alone before simultaneously measuring the DNA content and the electronic cell volume. If unavoidable effects on the DNA distribution are present, narrow forward-angle light scatter should be used as the cell size indicator during dual parameter measurements. Modifying instrument design by reversing electrode polarity might eliminate this problem.

Volume 28, Issue 4, pp. 330-334, 04/01/1980